How should a digital radiography sensor that is not heat tolerant be reprocessed?
작성자 : 관리자(info@kodexo.co.kr)  작성일 : 20.03.23   조회수 : 12
원문링크 https://www.cdc.gov/oralhealth/infectioncontrol/faqs/cleaning-disinfecting-environmental-surfaces.html
Digital radiography sensors are considered semicritical and should be protected with a Food and Drug Administration- (FDA-) cleared barrier to reduce contamination during use, followed by cleaning and heat sterilization or high-level disinfection between patients. If the item cannot tolerate these procedures then, at a minimum, it should be protected with an FDA-cleared barrier and cleaned and disinfected with an Environmental Protection Agency-registered hospital disinfectant with intermediate-level (i.e., tuberculocidal claim) activity, between patients. Because these items vary by manufacturer, refer to manufacturer instructions for reprocessing.
이전글 Can a dental office have carpeting and cloth furnishings in patient care areas?
다음글 Should contaminated radiology equipment be cleaned and disinfected, or protected with surface barrie